Content Determination of Loratadine in Human Plasma by HPLC
- VernacularTitle:HPLC法测定人血浆中氯雷他定的浓度
- Author:
Zifang TAN
- Publication Type:Journal Article
- Keywords:
Loratadine;
HPLC;
Plasma concentration
- From:
China Pharmacy
2007;0(30):-
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:To develop an HPLC method for the content determination of loratadine in human plasam.METHODS:The separation was performed on Kromasil C18(200 mm?4.6 mm,5 ?m) with mobile phase consisted of acetonitrile-water-0.5 mol?L-1 potassium phosphate-phosphoric acid(52:43:5:0.1) at flow rate of 1.0 mL?min-1.The detection wavelength was set at 275 nm and injection volume was 20 ?L.The column temperature was set at room temperature with diazepam as internal standard.RESULTS:The good linearity was obtained within 0.5~100 ?g?L-1(r=0.999 3).The detection limit was 0.5 ?g?L-1 and the recovery ranged during 99.71%~102.41%.The absolute recovery was within 80.72%~83.78%.The RSD of intra-day and inter-day were less than 11%.CONCLUSION:Established method could be used for the pharmacokinetics and bioavailability study of loratadine.