The Study of EABR and ECAP in Cochlear Implantation
- VernacularTitle:人工耳蜗植入者EABR与ECAP分析
- Author:
Xiaohua GE
;
Line WANG
;
Daoxing ZHANG
- Publication Type:Journal Article
- Keywords:
Electrically evoked auditory brainstem responses (EABR);
Electrically evoked compound action potential(ECAP);
Neural response telemetry (NRT)
- From:
Journal of Audiology and Speech Pathology
1997;0(04):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the threshold and latency of patients receiving cochlear implantation using electrically evoked auditory brainstem responses (EABR), and to evaluate the significance of EABR applied to those patients.Methods The EABR and ECAP were recorded in 14 subjects who were operated for Nucleus24 cochlear implant.They were assigned to Group A and Group B according to the results of ECAP. The threshholds of EABR, the wave latency of III and V and inter-wave latency of III-V of EABR were compared and analyzed between Group A and Group B.Results The threshholds of EABR were higher in Group B than in Group A on electrode No 22 when the pulse width was 25,50,75,100,125 ?s,respectively. There were significant differences between Group A and Group B(P0.05).Conclusion The threshholds of EABR were lower in patients whose ECAP findings were positive. The latencies of III and V,and inter-wave latency of III–V of EABR had no significant difference according to the positive findings of ECAP.