Content Determination of Itraconazole and Its Metabolit Hydroxy-itraconazole in Rabbit Plasma by HPLC-Fluorometric Methods
- VernacularTitle:HPLC-荧光法测定兔血浆中伊曲康唑及其代谢产物羟基伊曲康唑的浓度
- Author:
Wanghua YANG
;
Juan LI
;
Zicheng YU
- Publication Type:Journal Article
- Keywords:
Itraconazole;
Hydroxy-itraconazole;
Plasmas concentration;
HPLCfluorometric methods
- From:
China Pharmacy
2005;0(19):-
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE: To establish a high-performance liquid chromatographic(HPLC)-fluorometric method for the concentration determination of itraconazole (ITZ) and its metabolite hydroxy-itraconazole(HITZ) in rabbit plasma simultaneously.METHODS: Separation was achieved with Agilent Eclipse C18,the mobile phase consisted of acetonitrile(adjusted to pH=2.5 with 85% phosphoric acid)-water(63∶37),and the flow rate was 1.0 mL?min-1.The excitation and emission wavelengths were 260 nm and 365 nm.The analysis was performed at room temperature..RESULTS: The linear concentration range of ITZ and HITZ were 7.4~296 0 ng?mL-1(r=0.999 1)and 5.4~216 0 ng?mL-1(r=0.999 3), respectively.The limit of detection was 0.518 and 0.318 ng?mL-1,respectively.The extraction and method recovery rate of ITZ and HITZ were more than 80%.The intra-day and inter-day RSD of ITZ and HITZ were less than 4.47%. CONCLUSIONS: The method is simple,sensitive,accurate and reliable for monitoring ITZ and HITZ in rabbit plasma and pharmacokinetic studies.