Determination of Mycophenolic Acid in Kidney Transplantation Patients by HPLC
- VernacularTitle:HPLC法测定肾移植患者血浆中霉酚酸的浓度
- Author:
Lika YE
;
Zhihong XIE
;
Ruolun WANG
;
Xin XU
;
Weiting ZHANG
- Publication Type:Journal Article
- Keywords:
Mycophenolic acid;
HPLC;
Kidney transplantation patients;
Plasma drug concentration;
Multidoses;
Pharmacokinetics
- From:
China Pharmacy
2005;0(17):-
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:To determine the plasma mycophenolic acid concentration by HPLC, and study the multidoses pharmacokinetics character of mycophenolic acid in Chinese kidney transplantation patients. METHODS: The samples were precipitated with acetonitrile before injection. Diamonsil C18 column was used. The mobile phase consisted of acetonitrile-10 mmol?L-1 KH2PO4 (5∶6) at a flow rate of 1.1 mL?min-1. The detection wavelength was 254 nm and the column temperature was 40 ℃. This method was used to determine the multidoses pharmacokinetics in 12 kidney transplantation patients. RESULTS: MPA was well-separated from internal standard in chromatography, and endogenous foreign substance in plasma had no interference on the determination. The liner range for MPA was 0.38~59.00 ?g?mL-1,and the lowest detectable concentration of MPA was 0.38 ?g?mL-1. The recovery rate stood at 89.32%~97.63%; Both intra-day and inter-day RSD were less than 8.00%. Significant individual difference was noted among the patients treated with MMF in pharmacokinetic results, which was in line with the literature. CONCLUSION: This method is accurate and simple and applicable for the pharmacokinetics study of mycophenolic acid.