Determination of Pyridine in Potassium Dehydroandrograpolide Succinate by CGC
- VernacularTitle:毛细管气相色谱法测定穿琥宁中吡啶残留量
- Author:
Hongying MA
- Publication Type:Journal Article
- Keywords:
CGC;
Potassium dehydroandrograpolide succinate;
Pyridine;
Residual level
- From:
China Pharmacy
2005;0(15):-
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE: To establish a CGC method for the determination of pyridine in Potassium Dehydroandrograpol-ide Succinate. METHODS: The determination was performed on HP-INNOWAX capillary with column temperature at 100℃. The FID detector was used and the carrier gas was nitrogen. RESULTS:The linear range of Pyridine was 2.46~39.3?g?mL-1 and the average recovery was 98.07%,RSD=4.00(n=9).Its detection limit was 0.2?g?mL-1.CONCLUSION: The method is accurate and rapid, and applicable for the determination of the residual level of Pyridine in Potassium dehydroandrograpolide succinate.