Application of Active Fixation Lead in Patients with Right Ventricular Outflow Septum Pacing
- VernacularTitle:主动固定螺旋电极在右室流出道间隔部起搏中的应用体会
- Author:
Siming TAO
;
Ronghua ZHANG
;
Yi LI
;
Yunfei HONG
;
Feng YANG
;
Yun LU
- Publication Type:Journal Article
- Keywords:
Cardiology;
Active fixation lead;
Right ventricular septum
- From:
Journal of Kunming Medical University
1990;0(02):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the feasibility and methodology of active fixation lead on patients under right ventricular outflow tract septum(RVOTS) pacing.Methods Fifty DDD pacemaker patients were enrolled 31 male,23 female,50~86 years old,mean age 67.7?8.6.Ventricular active fixation lead was implanted in the right ventricular apex(RVA) and RVOTS successively and pacing parameter was tested.Results The success rate of RVOTS active fixation lead implantation was 98.15%.Mean lead threshold was 0.73?0.12 V.Pacing QRS duration show a significant difference between RVOTS pacing and RVA pacing,130.45?18.24 and 153.11?20.10,respectively(P