The confirmed inspection and analysis of HCV Different Region ELISA KIT to anti-HCV near S/CO sample
- VernacularTitle:丙型肝炎病毒分片段试剂对抗-HCV临界值附近标本的确认价值
- Author:
Xiaoyan XU
;
Ye FANG
;
Wanjun ZHENG
- Publication Type:Journal Article
- Keywords:
Hepatitis C antibodies;
Enzyme-linked immunosorbent assay;
Indicators and reagents;
Reference values
- From:
International Journal of Laboratory Medicine
2006;0(09):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To discuss the confirmed value of HCV Different Region ELISA KIT (ANTI-HCV-C, ANTI-HCV-NS3, ANTI-HCV-NS4, ANTI-HCV-NS5) to anti-HCV CO sample.Methods To inspect the general antibody of anti-HCV with ELISA, which is commonly used in clinic.The serum was separated from the suspicious samples, whose results were equal to S/CO or in the state of grey district ( CO?10%), and stored in the refrigerator in -29℃ low temperature and concenratedly inspected by the HCV Different Region ELISA KIT on the Triturus,the enayme immuno-reactions automatic analyzer. We totally collected 44 suspicious samples,among which there were 24 anti-HCV positive,20 anti-HCV CO?10% negative. Meanwhile we collected 30 samples whose S/CO of general antiboby was more than 5.0 and 20 ones which were healthy and could be compared with.Results In the 44 suspicious cases, whose S/CO was in the state of grey disrtict (CO?10%), the HCV Different Region Kit confirmed positive was 22.7% (10/44),suspicious 6.8% (3/44), negative 70.5 % (31/44).The false positive rate of positive anti-HCV CO?10% was 54.2%(13/24) and the false negative rate of negative anti-HCV CO?10% was 5.0%(1/20). In the 30 cases,whose S/CO of general antibody was more than 5.0,the HCV Different Region Kit confirmed positive was 80.0%(24/30),suspicious 13.3%(4/30), negative 67%(2/30). In 20 cases,which were from healthy people, the HCV Different Region Kit confirmed totally negative.Conclusion Using HCV Different Region Kit is a better evidence laboratory,especially for the uncertain samples,that is,it is very important for the "grey district"(CO?10%) samples to be further confirmed.It can reduce the false positive and false negative result which may emerge when anti HCV EIA inspects the dangerous groups and improve the accuracy of clinical HCV diagnosis.To blood station, it has certain practical value for the blood not only can be safely used, but also avoided wasting. The operation is simple and convenient, easily be spread and applied.