Determination of the Propofol Concentration of Plasma in Children Undergoing Cardiopulmonary Bypass by RP-HPLC
- VernacularTitle:RP-HPLC法测定儿童心肺体外循环后异丙酚的血浆浓度
- Author:
Hongbin GU
;
Yu CHEN
- Publication Type:Journal Article
- Keywords:
High pressure liquid chromatography;
Propofol;
Children;
Cardiopulmonary bypass;
Plasma
- From:
Journal of Chinese Physician
2002;0(S1):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To establish a HPLC method to detect the propofol concentration of plasma in children undergoing cardiopulmonary bypass. Methods RP-HPLC with fluorescence detector was used in the study. Plasma samples were precipitated by the thymol-methanol solution, and centrifugated. 20?l of supernatant fluid was injected for analysis and separated by Dikma Diamonsil C18 column. The mobile phase was composed of acetonitril:water =56:44(V/V), flo rate was 1.0ml/min. Fluorescence detective wave lengths were ?ex= 276nm, ?em=312nm. HPLC machine was Agilent1100 series. Main column was Diamonsil C18(150?4.6nm,5um),guard column was Phenomenex Security(4mm?3mm). Results There was a good linear relationship within the range of 0.02~8.0mg/L(r=0.9993, n =9),the limit of detection was 5?g/L,average recovery was 96.99%,intra RSD(relative standard deviation )was between0.810% and 5.73%,inter RSD was between 0.932% and 7.337%. Conclusions The method is sensitive, simple, rapid, good linear relationship and precise. The experiment can be used for determination of propofol concentration in child undergoing cardiopulmonary bypass.