The difference between target and measured concentration of remifentanil administered by target-controlled infusion: an evaluation of the performance of a new type-Ⅰ TCI system
- VernacularTitle:瑞芬太尼靶控输注时靶浓度和实测血药浓度的差异-靶控输注系统的性能评价
- Author:
Jiaqiang ZHANG
;
Yuying XING
;
Wenli DU
- Publication Type:Journal Article
- Keywords:
Piperidines;
Drug delivery system;
Evaluation studies
- From:
Chinese Journal of Anesthesiology
1996;0(09):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To determine the difference between target and measured concentration of remifentanil given by target-controlled infusion (TCI) and evaluate the performance of a new type Ⅰ TCI system for Chinese. Methods Thirty-six ASA Ⅰ or Ⅱ patients aged 40-60 yr weighing 50-70 kg undergoing elective lung resection were randomly divided into 2 groups according to target remifentanil concentration: group Ⅰ 6 ng ? ml-1 and group Ⅱ 8 ng?ml-1. The patients were premedicated with intramuscular midazolam 0.05 mg?kg-1 and atropine 0.5 mg. Anesthesia was induced with remifentanil and propofol both given by TCI. The target concentration of propofol (effect-site concentration) was set at 3 ?g?ml-1 and remifentanil (plasma concentration) at 6 or 8 ng? ml-1. When the patients lost consciousness, vecuronium 0.1 mg?kg-1 was given i. v. to facilitate intubation. The patients were mechanically ventilated and PETCO2 was maintained at 30-40 mm Hg. Anesthesia was maintained with TCI of propofol and remifentanil and intermittent i. v. boluses of vecuronium. Target plasma concentration of remifentanil remained unchanged during anesthesia. BIS value was maintained at 45-55 by modifying target propofol concentration. Arterial blood samples were taken before and 5, 10, 20, 40, 60, 90 and 120 min after TCI remifentanil was started for determination of blood remifentanil concentration by high performance liquid chromatography.The performance error (PE) was determined for each measured blood remifentanil concentration. The performance in the population was determined by median absolute performance error (MDAPE), median performance error (MDPE) and the wobble (the median absolute deviation of each PE from the MDPE). Results The measured concentrations (Cm) of remifentanil were significantly lower than the target plasma concentration (Cp) at5, 10, 20 min of TCI in both groups ( P