Influence of Changing the Parameters on the Wave of NRT
- VernacularTitle:参数改变对神经反应遥测波形的影响
- Author:
Xueqing CHEN
;
Shuo WANG
;
Demin HAN
- Publication Type:Journal Article
- Keywords:
Cochlear implant Electrically evoked compound action potential(ECAP) Neural response telemetry(NRT)
- From:
Journal of Audiology and Speech Pathology
1998;0(02):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To inevestigate the changes of the responses by changing the parameters in the neural response tetemety(NRT).Methods In this project,twenty patients who were implanted in Nucleus C124M were measured the electrically evoked compound action potential(ECAP)using the NRT system.It was expected to observe the changes of the responses by changing the parameters in the NRT 2.04 software.Results It was found that the parameters that made great influences to the responses were current level,stimulate pulse width,gain,delay and masker advance.Conclusion The parameters that made great influences to the record time were stimulate rate and number of sweeps.