Kinetic Analysis of Serum Creatinine
- VernacularTitle:血清肌酐的超微量速率法分析
- Author:
Yuming WANG
- Publication Type:Journal Article
- Keywords:
Serum creatinine;
Kinetic analysis;
Jaffe reaction;
Centrifugal analyzer
- From:
Journal of Kunming Medical University
1989;0(01):-
- CountryChina
- Language:Chinese
-
Abstract:
In this paper, we report Kinetic method was used to assess the concentration of human serum creatinine with MONARH 661 auto-centrifugal Analyzer. This method required small sample volumes and it was, rapid, convenient and reliable enough to be useful for routine de- termination. The standard curve of serum creatinine concentration was below 1768.0?mol/L. The recovery rate was 96.9-102%, within-run coefficient of variation was 2.86% and between-run was 4.22%. The major advanta ge of the method is that it can effectively avoid the negative interference from glucose, protein and bilirubin, etc. The reagent compares favorably with I Lco. reagent(X), Y = 0.99X-0.45,r=0.9995.