Determination of Glycemia-A comparision between Glucose-Oxidase and Hexokinase
- VernacularTitle:血液葡萄糖测定——葡萄糖氧化酶与已糖激酶之实验比较
- Author:
Mingrong YANG
- Publication Type:Journal Article
- Keywords:
Glucose oxidase;
hexokinase;
Qlucose;
dehydrogenase
- From:
Journal of Kunming Medical University
1989;0(01):-
- CountryChina
- Language:Chinese
-
Abstract:
Domestic Glucose-Oxidase (GOD) has been compared with imported Hexokinase (HK) in the test of glucose of human plasma. The result indicates that both are similar in accuracy and precision according to the referential method of HK. The intra-and inter-batches CV of GOD are 1.55% and 1.06% respectively. Its recovery rate is 96.7~105.2% while the concen- tration of glycemia is 2.8~28.0mmol/L, which indicates a good linear correlation. In the comparision of Trace-HK, Corning-HK, 1L-HK and corning-GOD, the correlation coefficients, r, the order is as follows: 0.999, 0.997, 0.996, and 0.996. 141 healthy people have been examined fer glycemia. Their normal values are 3.68~6.00mmol / L.