Analysis of the Sysmex XE 5000 blood cell analyzer for the analysis of abnormal lymphocytes and basophils
10.3760/cma.j.issn.1008-6706.2017.10.031
- VernacularTitle:Sysmex XE-5000血细胞分析仪对异型淋巴细胞和嗜碱性粒细胞的报警提示分析
- Author:
Xiaoqin HUO
- Keywords:
Hematology analyzer;
Abnormal lymphocyte;
Basophil;
Autoanalysis
- From:
Chinese Journal of Primary Medicine and Pharmacy
2017;24(10):1560-1562
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the value of Sysmex XE 5000 blood cell analyzer in alarm of atypical lymphocytes and basophil.Methods 198 samples of Sysmex XE 5000 blood cell analyzer and suggested that atypical lymphocytes and basophils, 915 copies of the instrument tip shaped specimens.Single lymphocyte increased blood smear microscopy for examination under a microscope to detect atypical lymphocytes as the gold standard and to verify the alarm not exactly.Results The microscopic examination results as the gold standard, and suggested that atypical lymphocytes and basophilia specimens after reinspection, eosinophilic granulocyte alkaline or positive value of 1.8%,in the normal range, the positive rate of atypical lymphocytes (129/198) was 65.2%,but the instrument alone atypical lymphocytes alarm specimens the positive rate of atypical lymphocytes after review of 72.4% microscope (663/915),the former was lower than the latter with significant difference (x2=4.521,P<0.05).Conclusion Sysmex XE 5000 blood cell analyzer at the same time of atypical lymphocyte and basophil alarm, two detection levels of interference, combined with microscope examination can improve detection accuracy, reduce the misdiagnosis rate, and provide reliable data for clinical treatment.