Non-invasive method of blood hemoglobin measurement based on time-resolved transmission spectroscopy
10.3760/cma.j.issn.1673-4181.2016.06.005
- VernacularTitle:基于时间分辨透射光谱的血液血红蛋白无创检测方法
- Author:
Yuan YUAN
;
He HUANG
;
Jichun YANG
;
Meixiu SUN
- Keywords:
Time-resolved transmission spectroscopy;
Blood hemoglobin;
Non-invasive measurement
- From:
International Journal of Biomedical Engineering
2016;39(6):341-344
- CountryChina
- Language:Chinese
-
Abstract:
Objective To theoretically study the non-invasive measurement method of blood hemoglobin in vivo based on the time-resolved approach combined with occlusion spectroscopy.Methods The dynamic dual wavelength time-resolved transmission was analyzed based on the artificial blood flow kinetics condition on a human finger model.The sensitivity of hemoglobin measurement was improved by Laplace transforming of time-domain data.The method was validated using hemoglobin with a mass concentration range of 6~16 g/dl.Results The simulation results showed that compared with the continuous wave method,the time-resolved method could provide higher detection sensitivity using Laplace transform parameter p =5 × 1010 s-1.Conclusions The sensitivity of hemoglobin measurement can be enhanced when early arriving photons are emphasized by Laplace transformingthe time-resolved transmission with a small positive parameter.