Application research of contrast material for coronary CT angiography by a combination of iterative reconstruction and double inj ection technology
10.3969/j.issn.1002-1671.2016.12.033
- VernacularTitle:双流技术调节对比剂浓度联合迭代重建在冠状动脉成像中的应用研究
- Author:
Junwu ZHANG
;
Xiong PANG
;
Zhenyu SHU
;
Zhen WANG
- Keywords:
contrast media;
computed tomography;
radiation dosage;
imaging procession
- From:
Journal of Practical Radiology
2016;32(12):1945-1950
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the feasibility of contrast material on image quality of coronary CT angiography (CCTA)by using double injection technology and iterative reconstruction.Methods 120 patients with suspected coronary heart disease who underwent CCTA were randomly divided into two groups.Then,60 patients with 30 kg/m2>BMI≥25 kg/m2 were averagely divided into A1 and B1 groups,and other 60 ones with BMI<25 kg/m2 were also averagely divided into A2 and B2 groups.The patients in group A were scanned with 120 kV tube output and iolromide 350 mg I/mL,and the images were reconstructed using filtered back projection (FBP).Meanwhile,the patients in group B were scanned with 80 kV or 100 kV tube output and 245 mg I/mL or 280 mg I/mL of contrast media (BMI≥25 kg/m2 )using double injection technology,and the images were reconstructed using adaptive iterative dose reduction 3D (AIDR 3D).The effective radiation dose (ED)of each patient was also calculated.CT attenuation of the main proximal vessels (ascending or descending aorta,pulmonary artery,RCA,LM,LAD and LCX)and the obj ective image quality (SD,CNR and SNR)were measured,and the subj ective evaluation was also assessed by an experienced radiologist.Results There was significant differences in mean image quality scores,SD,SNR and CNR of ascending aorta between group A1 and B1(P<0.05),but no difference in the SI of values (P>0.05).There was no significant difference in mean image quality scores between group A2 and B2 (P>0.05), however,there were significant differences in the SI,SD of values and SNR,CNR of ascending aorta (P<0.05).The values of BMI were negatively related to the values of CT in ascending aorta between two subgroups (r=-0.66 and -0.441,P<0.05).The values of weight were not related to the values of CT in ascending aorta (P>0.05).There was significant difference in effective dose between subgroups (P<0.05).Conclusion The method with a combination of iterative reconstruction and a contrast material of 245 mg I/mL using double inj ection technology can improve the contrast enhancement without impairing image quality.