Impact of movement frequency on gross tumor volume of moving tumors based on three-and four-dimensional CT scans
10.3760/cma.j.issn.1004-4221.2016.09.021
- VernacularTitle:3DCT和4DCT扫描下不同运动频率对运动中大体肿瘤体积的影响
- Author:
Dongping SHANG
;
Jianhong XING
;
Qiang ZHANG
;
Yong YIN
- Publication Type:Journal Article
- Keywords:
Respiratory movement;
Target volume;
Tomography,X-ray computed;
Body phantom
- From:
Chinese Journal of Radiation Oncology
2016;25(9):994-998
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the impact of movement frequency on gross tumor volume (GTV) of moving tumors and coordinate position of the central point based on three-and four-dimensional CT scans.Methods The respiratory motion platform from Modus and 8 phantoms with different shapes and volumes were used to simulate the movement of lung tumors.Three-and four-dimensional CT scans were performed at movement frequencies of 10,15,and 20 times/min.GTV (GTV10,GTV15,and GTV20) and IGTV (IGTV10,IGTV15,and IGTV20) were delineated,and the coordinate position of the central point was obtained.The Friedman test was performed for GTV10,GTV15,GTV20,IGTV10,IGTV15,IGTV20,and the coordinate position of the central point.Results GTV10,GTV15,and GTV20 at the three movement frequencies were 12.41±14.26 cm3,10.38±11.18 cm3,and 12.50±15.23 cm3,respectively (P=0.687),and the positional values were-8.2±96.2 mm,-8.6±96.1 mm,and-8.6±95.7 mm in x-axis (P=0.968),108.2±25.0 mm,110.4±22.5 mm,and 109.0±24.2 mm in y-axis (P=0.028),and 65.2±13.7 mm,65.4± 13.4 mm,and 65.4± 13.2 mm in z-axis (P =0.902).IGTV10,IGTV15,and IGTV20 at the three movement frequencies were 17.78± 19.42 cm3,17.43± 19.56 cm3,and 17.44± 18.80 cm3,respectively (P=0.417),and the positional values were-7.7±95.9 mm,-7.9±95.6 mm,and-7.9±95.1 mm in x-axis (P=0.325),109.4±24.5 mm,109.6±24.1 mm,and 109.2±24.3 mm in y-axis (P=0.525),and 65.5±13.3 mm,65.6±13.4 mm,and 65.5±13.3 mm in z-axis (P=0.093).Conclusions During simulated positioning of thoracic tumors,respiratory movement frequency has no significant impact on target volume established by four-dimensional CT scan.There are no significant differences in three-dimensional target volume established at different respiratory frequencies,but respiratory frequency has a significant impact on the position of the central point of the target volume in y-axis.