A exploration and application of a quality control method for urinary fluoride detection
10.3760/cma.j.issn.2095-4255.2016.10.018
- VernacularTitle:尿中氟化物批量检测的质量控制的应用
- Author:
Nianheng ZHANG
;
Dong AN
;
Dasheng LI
;
Yu YANG
;
Danya SHE
;
Yuehua PAN
;
Yuting HUANG
- Publication Type:Journal Article
- Keywords:
Urine;
Fluoride;
Quality control;
Application
- From:
Chinese Journal of Endemiology
2016;35(10):776-779
- CountryChina
- Language:Chinese
-
Abstract:
Objective To analyze the quality control measures on urinary fluoride testing in the implementation,and to provide experience in quality control for testing activities.Method According to the Determination of Fluoride in Urine-Ion Selective Electrode Method (WS/T 89-1996),focusing on ways to complete testing all the samples in a short time,implementing internal quality control measures in the testing before,during and after the implementation by different ways as blank experiment,standard substance detection,personnel parallel experiment,paralleled detection,instrument comparison and sample test repeat,were carried out.Results The test results of blank experiment were lower than the lowest detection limit;the test results of standard material were within the scope of the standard reference,there was no significant difference between the mean value and the reference value (t =0.01,0.00,0.02,all P > 0.05),|Z| values were all less than 1;the results of personnel parallel experiment,paralleled detection,instrument comparison test results and sample test repeat were not significantly different statistically (all P > 0.05);the results of instrument comparison and sample test repeat were not significantly different statistically (t =0.129,0.034,all P > 0.05).A total of 9 123 urine samples were tested,the geometric mean of urinery fluoride was 0.58 mg/L.Conclusion Different quality control measures should be implemented through all the testing,control personnel errors,instruments,and reagents,etc,to ensure the quality of the entire test results.