Simultaneous Determination of the Contents of 9 Metal Elements in Ganmaoling Granule by ICP-MS
	    		
		   		
		   			 
		   		
	    	
    	 
    	10.6039/j.issn.1001-0408.2015.30.27
   		
        
        	
        		- VernacularTitle:ICP-MS法同时测定感冒灵颗粒中9种金属元素的含量Δ
- Author:
	        		
		        		
		        		
			        		Xin SONG
			        		
			        		;
		        		
		        		
		        		
			        		Xueyu HANG
			        		
			        		;
		        		
		        		
		        		
			        		Lu WANG
			        		
			        		;
		        		
		        		
		        		
			        		Xiaoqing FENG
			        		
			        		;
		        		
		        		
		        		
			        		Qin WANG
			        		
			        		;
		        		
		        		
		        		
			        		Rui XU
			        		
			        		
		        		
		        		
		        		
 
			        		
			        		
		        		 
- Publication Type:Journal Article
- Keywords:
        			
	        			
	        				
	        				
			        		
				        		ICP-MS;
			        		
			        		
			        		
				        		Ganmaoling granule;
			        		
			        		
			        		
				        		Metal element;
			        		
			        		
			        		
				        		Microwave digestion
			        		
			        		
	        			
        			
        		
- From:
	            		
	            			China Pharmacy
	            		
	            		 2015;(30):4252-4254
	            	
            	
- CountryChina
- Language:Chinese
- 
		        	Abstract:
			       	
			       		
				        
				        	OBJECTIVE:To establish a method for simultaneous determination of the contents of Cu,Pb,As,Cr,Cd,Ba, Mn,Sb and Hg in Ganmaoling granule. METHODS:ICP-MS was conducted. The sample was handled by nitric acid microwave di-gestion system,the mass concnetration of 9 metal elements were determined by ICP-MS with Ge,In,Bi as internal standard and standard substance of test element standards. RF power was 1 100 W,sampling depth was 8 mm,carrier gas was argon(Ar)with high purity and flow rate was 1.0 L/min. RESULTS:All 9 elements had good linear range between mass concentration and ion peaks (r=0.999 1-0.999 9),detection limits were in the range of 0.3-6.0μg/L;RSDs of precision,stability and reproducibility tests were no more than 6.0%;average recoveries were in the range of 80.0%-113.5%,and RSDs were in 1.0%-4.5%. CONCLUSIONS:The method is simple,rapid and accurate,and can be used for the determination of 9 metal elements in Ganmaoling granule.