A methodologic study of quantitative analysis of contrast enhanced ultrasound in liver
10.3760/cma.j.issn.1004-4477.2016.05.011
- VernacularTitle:肝脏超声造影定量分析临床应用的方法学研究
- Author:
Chen YANG
;
Chanjuan PENG
;
Kaiyuan SHI
;
Na FENG
;
Weiyun PAN
;
Liyu CHEN
;
Dong XU
- Publication Type:Journal Article
- Keywords:
Contrast-enhanced ultrasound;
Liver;
Time-intensity curve;
Quantitative analysis
- From:
Chinese Journal of Ultrasonography
2016;25(5):400-404
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the effects of contrast dose and region of interest (ROI) depth on quantitative analysis of liver by contrast-enhanced ultrasound (CEUS) during clinical application.Methods After bolus injection of contrast agent,the change of quantitative parameters [including echo mean(EM),rise time(RT),peak intensity(PI),mean transit time(MTT),area under the curve(AUC),time from peak to one half(TPH),wash in slope(WIS),time to peak(TTP)] of time-intensity curves were analyzed based on groups from different doses (1.0 ml and 1.6 ml) and different depth (<30 mm,30-60 mm,and ≥60 mm).Results MTT and TPH were increased with dose increasing from 1.0 ml to 1.6 ml (P<0.05).With the dose 1.0 ml,TPH,WIS,PI,AUC and MTT showed significant difference when the depth of the ROI changes (P <0.05),with the depth increased,TPH,WIS,PI,and AUC all decreases and MTT increases.For all the other parameters,no significant changes were found (P >0.05).Conclusions CEUS and its imaging process can directly influence the accuracy of the parameters from the quantitative analysis.Standardization of contrast agent with predefined dose and depth can potentially facilitate future clinical studies in liver CEUS.