Evaluation of point-of-care testing for D-dimer in elderly patients
10.3760/cma.j.issn.1671-0282.2016.08.017
- VernacularTitle:老年患者应用D-二聚体床旁测定的性能评价
- Author:
Zeng ZENG
;
Hongmei JIAO
;
Chenxue QU
;
Dan SUN
;
Jiali DU
;
Dongfang WEI
;
Meng CHEN
- Publication Type:Journal Article
- Keywords:
D-dimer;
AQT90 FLEX immunoassay analyzer;
ACL TOP Coagulation analyzer
- From:
Chinese Journal of Emergency Medicine
2016;25(8):1049-1053
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the clinical performance of AQT90 FLEX,a novel time-resolved fluorescence based point-of-care test (POCT) for quantification of D-dimer in elderly patients.Methods The method from Quantitative D-dimer assay (WS/T 477-2015) for testing equipment performance was used as a reference to evaluate the clinical performance of AQT90 FLEX.The correlation was compared between testing results of D-dimer using the AQT90 immune-assay analyzer and those using the ACL TOP coagulation analyzer.Results At high concentration of D-dimer,the within-run precision coefficient of variation (CV)was 2.619%,and at low concentration of D-dimer,the within-run precision CV was 2.767%.The pollution-carrying rate was 0.12%.The measured data from AQT90 and ACL TOP had a correlation coefficient of r =0.9491 (P < 0.01).The equation of the line of best fit for D-dimer with which all AQT90 results can be adapted to the ACL TOP was:AQT90 =2.52 ACL TOP + 0.15.The number from the equation was slightly greater in female than that in male,and it was also increased in elderly.Conclusions The AQT90 FLEX had rational precision and linearity in determination of concentration.There was a high agreement between the testing results from AQT90 and those from ACL TOP.It was recommended to use a slope of 2.52 and an intercept of 0.15 to adjust the D-dimer values of the ACL TOP to the AQT90 FLEX assay systems.POCT for D-dimer by AQT90 FLEX raises feasibility for use in elderly patients.