Application of adaptive statistical iterative reconstruction in low dose temporal bone CT scan of children
10.3760/cma.j.issn.0254-5098.2016.07.014
- VernacularTitle:自适应统计迭代重建在儿童颞骨低剂量CT扫描中的应用研究
- Author:
Bin HE
;
Ying GUO
;
Ruijing YANG
;
Hongming LIU
;
Pengcheng WANG
- Publication Type:Journal Article
- Keywords:
Children;
Temporal bone;
Radiation dose;
Evaluation research
- From:
Chinese Journal of Radiological Medicine and Protection
2016;36(7):544-547
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate clinical use of adaptive statistical iterative reconstruction (ASiR) method in low-dose scan on children temporal bone by comparison of radiation dose delivered by GE Discovert HD CT 750 scan and GE Lightspeed VCT64 scan.Method Sixty patients with congenital deafness were divided into two groups according to gender,age (each N =30).Group a received low-dose CT scan on temporal bone by GE Discovery HDCT 750 (preset NI =12),and the original images were reconstructed with ASiR (weighted value =40%).Group B received CT scan by GE Lightspeed VCT 64 (preset NI =8),and the original images were reconstructed with filter back projection (FBP).Data were collected and analysed using the SPSS 18.0 software.Resultes The mean value of electric current of average tube and CTDIvol were significantly different between group A and B [(110.40 ± 21.72) mA vs.(168.56±24.36) mA,and (26.43 ±3.48) mGy vs.(39.66 ±4.17) mGy,respectively (t=-9.76,-3.31,P<0.05)].The noise index was (33.13 ±2.68) in group A and (33.79 ±2.93) in group B respectively,and the results had no statistical significance (P > 0.05).Subjective scores of the images were (4.06 ± 0.03) and (4.05 ± 0.03),with no statistical significance either (P > 0.05).Conclusions On a condition of achieving same image quality,CT scan program using GE Discovery HD CT 750 with 40% ASiR reconstruction and 4 units NI value enhancement could effectively reduce radiation dose on children's temporal bone compared to FBP method using GE Lightspeed VCT 64.