Research on quality testing for active spot scanning proton and heavy ion accelerator
10.3760/cma.j.issn.0254-5098.2016.08.013
- VernacularTitle:主动式点扫描质子重离子加速器质量检测研究
- Author:
Jinsheng CHENG
;
Jilong YUAN
;
Mingsheng LI
- Publication Type:Journal Article
- Keywords:
Proton and heavy ion accelerator;
Active spot scanning;
Quality test
- From:
Chinese Journal of Radiological Medicine and Protection
2016;36(8):621-625
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the quality testing of dose delivery system of the active spot scanning proton and heavy ion accelerator,in order to provide the reference for the quality control of related equipment.Methods In the four therapy rooms,both 0.6 cc chambers and Gafchromic EBT3 films were used,respectively,to test the accelerator for dose reproducibility,dose linearity,dose stability,depth dose distribution,beam scanning position deviation and radiation field uniformity in each therapy room.Results Dose reproducibility variation coefficients are all less than 1.5%,dose linearity's maximum deviations less than 2%,dose stability's deviations less than 2%,depth dose distribution stability within 2%,beam scanning position deviation less than 1 mm,consistency of irradiation field's deviation less than 2 mm,and flatness within ± 5%.Conclusions The indicators about quality testing for the active spot scanning proton and heavy ion accelerator are all in line with the requirements of IEC standards draft.