The solutions and common problems analysis on the application of syringe pump
10.3969/J.ISSN.1672-8270.2016.06.037
- VernacularTitle:微量注射泵应用常见问题及解决方案
- Author:
Zhenhuan DONG
;
Haiyan ZHANG
;
Tiejun LI
- Publication Type:Journal Article
- Keywords:
Trace injection pump;
Problem;
Treatment measures
- From:
China Medical Equipment
2016;13(6):115-117
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To study the trace injection pump in the application of the common problems and treatment measures.Methods: 246 cases of general surgery patients were used trace injection pump problems for retrospective analysis, in order to propose the improvements about engineering and management.Results: The incidence rate of non-standard operation procedures or the incidents are 6.1%, injection pump failure and infusion speed is 2.03%, mismatched injection pump attachments is 1.22% and phlebophlogosis is 0.41%. The non-standard operation procedures and imperfect trace injection pump management are the major problems. Therefore, some solutions are proposed, which includes strengthening technical training for medical staff, standard operation procedures, regular quality monitoring and healthy education to the patients.Conclusion: The operation procedures of trace injection pump should be standard and programmatic. It is of great significance to clinical application of the trace injection pump and to work properly by strengthening operation personnel quality, dealing with problems timely and correctly.