Three-dimensional finite element analysis of angled abutments in anterior maxilla implant restoration
10.7652/jdyxb201603021
- VernacularTitle:上前牙区角度基台应用的三维有限元研究
- Author:
Liangzhi DU
;
Zhe XU
;
Ping SHI
;
Juan CHAI
;
Xiaofeng CHANG
- Publication Type:Journal Article
- Keywords:
angled abutment;
implant;
finite element method (FEM);
anterior maxilla
- From:
Journal of Xi'an Jiaotong University(Medical Sciences)
2016;37(3):403-407
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the effects of angled abutments on the anterior maxilla implant restoration. Methods We analyzed the biomechanical properties of implants of different sizes (Φ3.5 mm,4.0 mm and 4.5 mm in diameter;L11.5 mm and L13 mm in length)after connecting different angled abutments (0°,10°,20°,and 30°) using finite element method.Results The stresses and strains of loading parts of restorations increased and their distribution became more concentrated as the angle of abutment increased.Cortical bone of Φ3 .5 implants with smaller angle (10°or less)andΦ4.0 implants with abutments had the risk of overpassing the bone elastic threshold when the angle approached 30°.However,the cortical bone elastic deformation was within a safe range at all angles inΦ4.5 group.Conclusion We should consider the diameter of the implant when selecting angled abutments.The angled abutments are not suitable for small diameter implants.The bite force should be under control when needed. The larger angled abutments can be applied in the standard and major diameter implants and it is necessary to avoid occlusal overloading.