Clinical study on application of non-invasive and dynamic cerebral edema monitor during perioperative period of cardiopulmonary bypass in infantile congenital heart disease
10.3969/j.issn.1671-8348.2016.12.015
- VernacularTitle:无创脑水肿动态监护仪在小儿先天性心脏病体外循环围术期的临床研究
- Author:
Jiahui DING
;
Zhonglin WANG
- Publication Type:Journal Article
- Keywords:
extracorporeal circulation;
brain damage;
impedance coefficient
- From:
Chongqing Medicine
2016;45(12):1632-1634
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the influence of cardiopulmonary bypass(CPB) on brain injury in the infants with compli‐cated congenital heart disease(CHD) and simple CHD by analyzing and monitoring the perioperative cerebral electric impedance co‐efficient ,serum S‐100βprotein and neuron specific enolase (NSE) .Methods Forty infants with CHD were randomly selected and divided into two groups:the complicated group(n=20 ,13 cases of tertralogy of fallot ,7 cases of transposition of great arteries) and the simple group(n=20 ,8 cases of atrial septal defect ,12 cases of ventricular septal defect) .Blood samples were taken and the cere‐bral electric impedance coefficients before operation(T1 ) ,at 10 min after CPB(T2 ) ,immediately after aortic clamping(T3 ) ,aortic o‐pening(T4 ) ,end of CPB(T5 ) ,at 5 h(T6 ) and 24 h (T7 ) after CPB were measured by the BORN‐BE noninvasive brain edema moni‐tor .Serum levels of S‐100βprotein and NSE were detected by ELISA .Results There was no statistically significant difference in the cerebral electric impedance coefficient ,S‐100βprotein and NSE before surgery(P>0 .05) .The cerebral electric impedance coef‐ficient ,S‐100βprotein and NSE were significantly increased from the beginning to the end of CPB(P< 0 .05) ,and gradually de‐creased after CPB(P<0 .05) .The cerebral electric impedance coefficient ,S‐100βprotein and NSE were obviously higher in the com‐plicated group than those in the simple group at the same time points(P<0 .05) .The cerebral electric impedance coefficient was re‐spectively correlated with S‐100βprotein and NSE .Conclusion The influence of CPB on brain injury in complicating CHD infants is more serious than that in simple CHD infants .The BORN‐BE noninvasive brain edema monitor can be used during CPB periopera‐tive period in complicating CHD infants for conducting the real time monitoring in order to reduce brain damage .