Clinical observation of the marginal and internal fit of titanium crowns fabricated by a selective laser mel-ting technology
	    		
		   		
		   			 
		   		
	    	
    	 
    	10.3969/j.issn.1001-3733.2016.03.004
   		
        
        	
        		- VernacularTitle:选择性激光熔覆(SLM)纯钛全冠适合性的临床研究
- Author:
	        		
		        		
		        		
			        		Kenan CHEN
			        		
			        		;
		        		
		        		
		        		
			        		Jinghong MAO
			        		
			        		;
		        		
		        		
		        		
			        		Yuqi DANG
			        		
			        		;
		        		
		        		
		        		
			        		Bo GAO
			        		
			        		
		        		
		        		
		        		
 
			        		
			        		
		        		 
- Publication Type:Journal Article
- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Selective laser melting(SLM);
			        		
			        		
			        		
				        		Crowns;
			        		
			        		
			        		
				        		Fit;
			        		
			        		
			        		
				        		CAD/CAM;
			        		
			        		
			        		
				        		3D Printing;
			        		
			        		
			        		
				        		Clinical trial
			        		
			        		
	        			
        			
        		
- From:
	            		
	            			Journal of Practical Stomatology
	            		
	            		 2016;32(3):317-320
	            	
            	
- CountryChina
- Language:Chinese
- 
		        	Abstract:
			       	
			       		
				        
				        	Objective:To compare the marginal and internal fit of selective laser melting(SLM)titanium crowns with lost-wax cast (LW)titanium crowns.Methods:Titanium crowns of 10 subjects were fabricated by SLM and conventional LW respectively(n =10).The marginal and internal gaps of the crowns were recorded with silicon film using a replica technique.Each silicon film was cut into 2 parts and the thickness of silicon layer was measured at ×100 magnification using a stereomicroscope,the data of marginal gap (MG)and internal gap(IG)were statistically analysed by ANOVA and SPSS statistical package version 17.0.Results:The MG (μm)of the titanium crowns of SLMgroup and LMgroup were 90.67 ±14.7 and 94.77 ±21.9(P <0.05),IG were 213.73 ±90.4 and 217.00 ±97.7(P >0.05),respectively.Conclusion:The marginal fit of the SLMcrowns is better than that of the LW ones,and is significantly smaller than 120 μm of the clinical generally accepted standards.