Tools and methods of intensity-modulated radiation therapy dose verification
10.3760/cma.j.issn.1673-4181.2016.01.011
- VernacularTitle:调强放射治疗剂量验证工具与方法
- Author:
Jianfeng SUI
;
Liugang GAO
;
Xinye NI
- Publication Type:Journal Article
- Keywords:
Intensity-modulated radiation therapy;
High conformity;
Deviation;
Dose verification
- From:
International Journal of Biomedical Engineering
2016;39(1):49-53
- CountryChina
- Language:Chinese
-
Abstract:
Intensity-modulated radiation therapy is widely used in the treatment of tumor,and the dose distribution highly conforms to tumor target area in three-dimension.However,the factors such as complex beam,data error,algorithm error and machine error may cause large dose deviation in intensity-modulated radiation therapy,which may lead to unnecessary radiation accident.Therefore,standing on the patients' safety point of view,dose verification is usually performed before executing the treatment plan in order to ensure the safe implementation of the treatment plan and to avoid un-planned irradiation dose.Currently,there are many tools and methods of dose verification in clinic,including point dose verification tools like finger-shaped ionization chamber and thermoluminescence dosimeter;two-dimensional dose verification tools like Mapcheck,MatriXX and films;three-dimensional dose verification tools like ArcCHECK,Delta4 and the third-party software.These common clinical dose verification methods are reviewed in this paper.