Dosimetric comparison of stereotactic body radiation therapy plans for spine metastasis using flattening filter free and flattening filter modes
10.3760/cma.j.issn.0254-5098.2016.01.012
- VernacularTitle:有无均整器模式下脊柱转移瘤立体定向放疗计划的剂量学比较
- Author:
Tingting ZHUANG
;
Baihan LIN
;
Dongsheng LI
;
Lili WU
- Publication Type:Journal Article
- Keywords:
Spine metastasis;
Stereotactic body radiation therapy;
Flattening filter free;
Dosimetry
- From:
Chinese Journal of Radiological Medicine and Protection
2016;36(1):63-66
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the dosimetric difference of stereotactic body radiation therapy (SBRT) plans for spine metastasis using 6 MV X-ray flattening filter free (FFF) and flattening filter (FF) modes.Methods A total of nine previously treated patients with a total of twelve spine metastasis were retrospectively included and replanned using FFF-SBRT and FF-SBRT, respectively.The dose-volume histograms (DVH), target conformity index (CI), monitor unites (MUs) and treatment time of target volume, organs at risk and normal tissues were compared between the plans.Results Both FFF-SBRT and FF-SBRT met the clinical objectives.Dose distribution of target volume, organs at risk and normal tissues were similar.MUs of the FFF-SBRT were significantly higher than that of FF-SBRT (t =-5.20, P <0.0l), while the treatment time was almost half of the latter (t =17.27, P < 0.01).Conclusions Two plan modes are both clinically acceptable.FFF-SBRT plans are delivered in less time and with better efficiency.Trial registration Chinese clinical trial registry, ChiCTR-TRC-14004281.