Establishment of quality control evaluation system for 3D-ultrasound-based image-guided radiotherapy
10.3760/cma.j.issn.1004-4221.2016.03.016
- VernacularTitle:三维超声图像引导系统质控评价体系建立
- Author:
Xiaobo WEI
;
Qiang HU
;
Jinming MU
;
Guoping XU
;
Changhai CHENG
;
Qilin LI
;
Honglei PEI
- Publication Type:Journal Article
- Keywords:
Three-dimensional ultrasound;
Image-guided radiation therapy;
Quality control;
Phantom
- From:
Chinese Journal of Radiation Oncology
2016;(3):270-274
- CountryChina
- Language:Chinese
-
Abstract:
Objective To analyze the stability and accuracy of the equipment for three-dimensional ultrasound-based image-guided radiation therapy (3DUS-IGRT) in daily practice, and to provide a basis for clinical application of radiotherapy for soft tissue tumors.Methods A specific calibration phantom was used for continuous calibration and quality control of the 3DUS-IGRT equipment in a year.The method for daily quality control of ultrasound-guided equipment was explored, and its stability and accuracy were monitored.Results The phantom position errors in both Sim and Guide stations of the 3DUS-IGRT equipment were within 1 mm.Conclusions The 3DUS-IGRT equipment has a stable performance with the support of a complete set of stringent and accurate calibration and quality control, which provides a new image-guided method for precise radiotherapy for soft tissue tumors.