Optimization of Performance of Toroidal Ion Trap with Triangular Electrode by Theoretical Simulation
10.11895/j.issn.0253-3820.150900
- VernacularTitle:新型三角形电极圆环离子阱的理论模拟研究
- Author:
Haiyang YANG
;
Chongsheng XU
;
Lei YUE
;
Sudakov MIKHAIL
;
Yuanjiang PAN
;
Chuanfan DING
- Publication Type:Journal Article
- Keywords:
Toroidal ion trap;
Theoretical simulation;
Asymmetric electrodes;
Mass resolution;
High-order multipoles analysis
- From:
Chinese Journal of Analytical Chemistry
2016;(3):482-488
- CountryChina
- Language:Chinese
-
Abstract:
The toroidal ion trap is an ideal candidate for miniaturized ion trap because it has much higher ion trapping capacity than a standard quadrupole ion trap of equal trapping dimensions. A novel toroidal ion trap mass analyzer with triangular electrode which contained a filament end cap, a detector endcap, an inner ring and an outer ring was reported. After designing and optimizing the electrodes by theoretical simulations, we found that the asymmetric triangle electrodes could reduce the affection from toroidal shape and improve the ion ejection rate and the mass resolution of the ion trap. The best design of the toroidal ion trap with a mass resolution of 1486 at m/z 609 was obtained.