The working principle of the dialysate heating and mixture ratio system and the effect of dialysate temperature on renal fibrosis in dialysis patients with end-stage renal
10.3969/J.ISSN.1672-8270.2016.03.004
- VernacularTitle:血液透析装置透析液加热配比系统技术分析及透析液温度对肾纤维化透析患者的影响研究
- Author:
Jiawei YAN
;
Liqun HE
- Publication Type:Journal Article
- Keywords:
Hemodialysis device;
Dialysate;
Temperature;
Mixture ratio;
Temperature monitoring
- From:
China Medical Equipment
2016;13(3):18-21
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To explore the significance of blood dialysis device for real-time monitoring of dialysate temperature.Methods: Nikkiso DBB-27-type hemodialysis device was selected as a subject for discussion, analyzed and discussed the working principle, main structure and functional characteristics of the mixture ratio of dialysate and temperature monitor and combined with one case of temperature troubleshooting case.Results: Temperature monitoring was essential to ensure the stable running of dialysate mixture ratio and the temperature of the dialysate.Conclusion: The dialysate temperature has an important effect on renal fibrosis in end-stage hemodialysis patients, it is necessary to monitor the dialysate temperature and mixture ratio. Dialysis device heating degas proportioning system through surveillance system linked to the dialysis unit normal, stable operation, provide safe and reliable dialysis treatment for patients.