Study of sensitivity for detecting small positional deviation of MLC by 3D Detector array Delta-4
10.3760/cma.j.issn.1004-4221.2016.02.018
- VernacularTitle:Delta-4探测MLC细小位置偏差敏感性研究
- Author:
Xinye NI
;
Liugang GAO
;
Tao LIN
- Publication Type:Journal Article
- Keywords:
Multi-leave collimators;
Dose;
Position;
Deviation
- From:
Chinese Journal of Radiation Oncology
2016;25(2):168-171
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the detective sensitivity for position of multi-leave collimators ( MLC) using Delta-4. Methods First,the small positional deviation of MLC was simulated and measured using the linac (Varian,Trubeam) equipped with EPID.Then,two beam fields 2. 0 cm (x)×6. 0 cm (y), 7. 0 cm (x)×6. 0 cm (y) were designed,the x1 and x2 of MLC were expanded 0. 1 mm,0. 2 mm,0. 3 mm... 0. 9 mm and 1. 0 mm,2. 0...5. 0 mm to external simultaneously,different parameters of 3 mm/3%,2. 5 mm/2. 5%,2 mm/2%, 1. 5 mm/1. 5% and 1 mm/1% were used in Gamma analysis to analyze the difference between dose distribution detected by Delta-4 and original dose distribution with unexpanded MLC position derived from TPS. Results For 2. 0 cm ( x) × 6. 0 cm ( y) beam field, the pass rate of original dose distribution was 100%,and that decreased to 95. 5% when x1 ,x2 of MLC were expanded 0. 3 mm to external, and decreased to 89. 4% when expanded 0. 5 mm at 2. 5 mm/2. 5% statistical standards. For 7. 0 cm ( x) × 6. 0 cm ( y) beam field,the pass rate of original dose distribution was 96. 5%,and that decreased to less than 95% when x2,x2 of MLC were expanded 0. 3 mm to external,and passing rate was above 90% when MLC expanded less than 0. 5 mm at 1. 5 mm/1. 5% statistical standards. Conclusions For MLC' s positional deviation in decimillimeter level,raise standards of Gamma analysis properly may improve the capability of Delta-4 for detecting small positional deviation,but it won' t detect all the positional deviation of MLC in decimillimeter level. For different size of beam field,it is proposed to use different analytical standards for Delta-4.