Evaluation verification of Mindary BC5390 hematology analyzer
10.3969/j.issn.1673-4130.2016.08.013
- VernacularTitle:迈瑞BC5390血液分析仪实验室性能验证
- Author:
Yongqiong WEI
;
Wei SHEN
;
Chenggui LIU
;
Dengcheng CAO
;
Lian YANG
- Publication Type:Journal Article
- Keywords:
hematology analyzer;
performance qualification;
precision;
accuracy
- From:
International Journal of Laboratory Medicine
2016;37(8):1046-1048
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the performance verification of Mindary BC5390 hematology analyzer .Methods According to requirement of CLSI documents ,the precision ,accuracy ,linearity ,carryover were evaluated .Results The background counts all met the designed requirements of manufacturer .Precision ,accuracy ,linearity of the WBC ,RBC ,PLT ,Hb ,HCT ,MCV were good and the contamination carrying rates were low .Except for basophils ,the classification of leukocyte was well correlated with the artificial microscopy classification(r2 >0 .95) .Conclusion The BC5390 is a ideal automated hematology analyzer which has high precision and accuracy ,wide linearity ,low contamination carrying rate ,and it can be used for the blood samples analysis .