The application research of endoscopic equipment with power through detection of electrical prop-erties in sterilization and supply center
10.3760/cma.j.issn.1672-7088.2016.12.015
- VernacularTitle:带电源腔镜器械通电性能检测在消毒供应中心的应用性研究
- Author:
Yanling BI
- Publication Type:Journal Article
- Keywords:
Laparoscopes;
Surgical instruments;
Electric conductivity;
Sterilization and supply center;
Application research
- From:
Chinese Journal of Practical Nursing
2016;32(12):931-933
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the effects of the failure rate of laparoscopic apparatus with power through detecting the electrical properties before packaging it. Methods The laparoscopic apparatus with power used in the operation in January to June 2014 were randomly selected and divided into the control group and the observation group according to whether they underwent electricity performance detection. The control group did not receive testing of electrical properties before packing. The observation group tested the electrical conductivity of the single or double pole conductor, unipolar electric coagulation hook, bipolar electric coagulation forceps, the handle of electric knife by digital multimeter and the performance of the uterus rotary cutting machine by electric hysterectomy controller before packing. The difference of equipment in the number of failures, the type of failure, the number of cases and the satisfaction of medical personnel were observed between the two groups. Results In the observation group, 12 cases with unipolar and bipolar coagulation failure, 10 cases with electrocision handle failure, 2 cases with uterine rotary- cutting failure, 18 cases with adverse conductive or non-conductive, 6 cases with short circuit, which were significantly lower than those in the control group (36,25, 11, 48, and 26 cases), the differences were statistically significant (χ2=5.627-12.063, P<0.05). The satisfaction degree in the observation group was 94.74% (36/38), which was better than 76.32%(29/38) of the control group, the difference was statistically significant (Z=-2.267, P<0.05). Conclusions Testing the electric performance of laparoscopic apparatus with power before packing could reduce the incidence of intraoperative equipment failure rate, This method can provide safe and high quality surgical instruments.