Study of low kV in reducing pulmonary CT imaging contrast dose and radiation dose
10.3969/j.issn.1002-1671.2016.03.029
- VernacularTitle:低管电压在降低肺动脉CT成像对比剂量及辐射剂量的研究
- Author:
Jiquan SUN
;
Wenxin CHEN
;
Zhiqiang WU
;
Shichang TANG
;
Jingtao WU
- Publication Type:Journal Article
- Keywords:
computed tomography;
pulmonary artery;
iodine;
effectively absorbed dose
- From:
Journal of Practical Radiology
2016;32(3):437-440
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the iodine contrast agent dosage,radiation dose and image quality in CT pulmonary angiogra-phy (CTPA)with low tube voltage and high tube current in 64-slice spiral CT.Methods 60 patients with suspected pulmonary em-bolism in our hospital were randomly chosen and divided into three groups:Group Ⅰ:20 patients,using 120 kV,180 mA,contrast dose of 70 mL.Group Ⅱ:20 patients,using 100 kV,280 mA,contrast dose of 50 mL.Group Ⅲ:20 patients,using 80 kV,automat-ic tube current modulation techniques (300-500 mA),comparative dose of 30 mL.CT values and image noise of three pulmonary central areas and a segment branch were measured in each group.Objective indicators,subjective image quality assessment,CT dose index volume (CTDIvol),dose-length product (DLP)and effective absorbed dose (ED)were compared in three groups to evaluate the value of low kV,high mA with low-contrast dose in CTPA.Analysis of variance and t-test was used for data analysis.Results Compared with those of the standard method of CTPA,all pulmonary dry in two groups of low dose were well displayed.There was no statistical significance between each two groups in image quality score (P > 0.05),but was statistical significance in CT values, noise,SNR and CNR (P < 0.01).The CT radiation dose,CTDIvol and DLP of Group Ⅱ and Group Ⅲ were significantly lower than those of Group Ⅰ (P < 0.01).Conclusion Compared with traditional methods,low dose contrast agent injections under 80 kV could reduce the dose of contrast agent and the effect of hardening artifacts due to contrast agent in superior vena cava on right pul-monary artery.It is helpful to reduce the risk of contrast induced nephropathy and reduce patients’exposure to X-ray radiation.