The effect of postoperative central vault on the anterior chamber after implantation of implantable contact lens V4c
10.3969/j.issn.1671-8348.2015.36.014
- VernacularTitle:有晶体眼后房型人工晶体 V4c 植入术后拱高对前房结构的影响
- Author:
Jin YE
;
Changming YUAN
;
Jixue WANG
;
Hanping XIE
- Publication Type:Journal Article
- Keywords:
implantable contact lens;
anterior chamber angle;
anterior chamber volume;
postoperative central vault;
Pentacam
- From:
Chongqing Medicine
2015;(36):5092-5094
- CountryChina
- Language:Chinese
-
Abstract:
Objective To observe the effect of postoperative central vault on the anterior chamber after implantation of im-plantable contact lens V4c .Methods To carry out a retrospective study of 56 eyes with implantable contact lens V4c implantation in Chongqing Bright Eye Hospital from November 2 ,2014 to February 5 ,2015 ,examined anterior chamber angle before surgery ,the anterior chamber angular width (ACA) and anterior chamber volume (ACV) were measured by pentacam .the postoperative central vault and ACA and ACV during the first and third and sixth months after surgery were measured ,the pigment grade in anterior chamber after operation were recorded .The change of ACA and ACV of the 56 eyes was correlation analyzed .Results At first month after surgery ,the postoperative central vault was 250 ~ 765 μm ,the reduced amount of ACA was (13 .67 ± 1 .67)° ,the re-duced amount of ACV was (94 .23 ± 17 .02)mm3 .At the third month ,the postoperative central vault of the 55 eyes was 239 ~ 761μm ,the reduced amount of ACA was (13 .70 ± 1 .68)° ,The reduced amount of ACV was (93 .6 ± 17 .48)mm3 .At the sixth month , the postoperative central vault of the 51 eyes was 246 ~ 761 μm ,the reduced amount of ACA was (13 .60 ± 1 .69)° ,the reduced a-mount of ACV was(94 .29 ± 17 .80)mm3 .No case of pigment was found for grade Ⅱ .The statistical comparison of the amount of ACV and ACA decreased in different postoperative central vaults showed that the difference was statistically significant ,there were statistically significant differences in the amount of ACV and ACA between different central postoperative central vault(P< 0 .05) . Conclusion Anterior chamber angle was narrowed and anterior chamber volume decreased in the posterior chamber intraocular lensimplant-ation after implantable contact lens V4c implantation ,the postoperative central vault is related to the change of anterior chamber angle and ACV .Appropriate postoperative central vault has positive clinical significance for keeping anterior chamber angle open .