Clinical Follow-up Results in 33 Patients With Pacemaker Implantation With Home Monitoring System
10.3969/j.issn.1000-3614.2015.12.011
- VernacularTitle:植入带有家庭监测系统起搏器33例患者的临床随访结果
- Author:
Xudong GONG
;
Jun JIANG
- Publication Type:Journal Article
- Keywords:
Palemaker,artificial;
Deifbrillators,implantable;
Follow-up study
- From:
Chinese Circulation Journal
2015;(12):1178-1181
- CountryChina
- Language:Chinese
-
Abstract:
Objective: To study the clinical value of home monitoring system of pacemaker and implantable cardioverter-deifbrillators (ICD) in relevant patients.
Methods: A total of 33 patients with home monitoring system of pacemaker implantation or ICD treated in our hospital from 2011-02 to 2014-07 were retrospectively recruited. There were 28 patients with dual chamber pacemaker, 4 with ICD and 1 with cardiac resynchronization therapy pacemaker (CRTP). The patients included 23 male and 10 female at the age of (68.4 ± 9.2) years. The rate of successful data transmission, threshold changes and arrhythmia occurrence were analyzed.
Results: Up to 2014-10, a total 24516 messages were successfully transmitted with the success rate of 97.4%, the rate of no-message transmission was 2.6%. The home monitoring system fund 89 abnormal events including 73 disease related events (82.0%) and 16 system related events (18.0%).
Conclusion: Home monitoring system was a safe, reliable and effective monitoring equipment for cardiovascular implantable electronic devices; according to transmitted message, the parameters of devices could be adjusted in time for treating arrhythmia in relevant patients.