Cleaning and disinfection methods of medical tablet computers
10.3969/j.issn.1671-9638.2015.12.012
- VernacularTitle:医用小型平板电脑现场清洁消毒方法研究
- Author:
Huifen LI
;
Weihong ZHANG
;
Songqin LI
;
Suming ZHANG
;
Xiang ZHANG
;
Bo LIU
- Publication Type:Journal Article
- Keywords:
tablet computer;
cleaning;
disinfection;
adenosine triphosphate;
ATP;
biofluorescence
- From:
Chinese Journal of Infection Control
2015;(12):834-836
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the contamination of clinically used tablet computers,and compare the effec-tiveness of three cleaning and disinfection methods.Methods The front and back surfaces of tablet computers were wiped and detected by handheld adenosine triphosphate (ATP)fluorescence detector,the study was divided into baseline stage and cleaning and disinfection stage.During baseline stage,the tablet computers in use were directly wiped,during cleaning and disinfection stage,the tablet computers were wiped by normal saline-moistened gauze, alcohol-moistened gauze,and wet napkin.Results During baseline stage:20 tablet computers were wiped,the qualified rate was 0,the median of relative light unit (RLU)of ATP detection was 218.00.During cleaning and dis-infection stage,10 tablet computers were wiped in each cleaning and disinfection group,and the qualified rate of normal saline-moistened gauze,alcohol-moistened gauze,and wet napkin groups were 50.00%,0,and 60.00% re-spectively,the median RLU of ATP detection were 28.50,79.00,and 29.00 respectively.Except comparison be-tween saline-moistened gauze and wet napkin groups (P =0.97),multiple comparison of RLU of ATP detection amongthe other groups were significantly different (all P <0.001 ).Conclusion Contamination of medical tablet computers are serious,wet napkin wiping is an ideal method for cleaning and disinfection,but the frequency for cleaning and disinfection needs to be further studied.