The interference evaluation of hyperlipidemia and hyperbilirubinaemia to HbA1c measurement with IE-HPLC method
10.3969/j.issn.1673-4130.2015.17.016
- VernacularTitle:高血脂高胆红素对IE-H PLC法检测糖化血红蛋白干扰的评价
- Author:
Jinli XIAO
;
Xiuming ZHANG
;
Shengnan XU
;
Minghuan SUO
;
Quanzhong XU
;
Yaqiong CHEN
;
Jianyang WU
;
Man LI
;
Lijuan KAN
;
Dongmei WEN
- Publication Type:Journal Article
- Keywords:
diabetes;
glycated hemoglobin;
hyperlipidemia;
hyperbilirubinaemia
- From:
International Journal of Laboratory Medicine
2015;(17):2492-2494
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the interference of hyperlipidemia and hyperbilirubinaemia to HbA1c measurements by ion‐exchange high‐performance liquid chromatography(IE‐HPLC) method .Methods Fresh whole‐blood samples collected with EDTA‐K2 anticoagulant tubes were divided into four groups :control group(HbA1c<6 .2% ) ,diabetes group(HbA1c≥6 .2% ) ,hyperlipi‐demia group(TG 3 -20 mmol/L);hyperbilirubinaemis group (TBIL 21 -549 μmol/L) .HbA1c of these samples were measured with affinity chromatography(AC‐HPLC) and IE‐HPLC respectively .Results When HbA1c≤18 .7% ,r=0 .993 ;95% confidence interval(CI) of HbA1c results by using IE‐HPLC method was -0 .71 -0 .89 ;coefficient of variation was -5 .8% -6 .8% ;P=0 .198 and the difference was not statistically significant .When HbA1c< 16 .3% ,r= 0 .997;95% CI of HbA1c results with IE‐HPLC method is -0 .31-0 .67;coefficient of variation was -5 .8% -4 .3% .P=0 .000 and the difference was statistically signifi‐cant .No interference was detectded with the results ;When HbA1c was 16 .3% -18 .7% ,positive bias was observed with the re‐sults .When TG≤20 .78 mmol/L ,r=0 .995;95% CI of HbA1c results with IE‐HPLC method was -0 .26-0 .50 ;coefficient of var‐iation was -5 .5% -5 .8% .P=0 .000 and the difference was statistically significant .No interference was detectded with the re‐sults;When TBIL≤549 .3 μmol/L ,r=0 .990 ;95% CI of HbA1c results with IE‐HPLC method was -0 .08 -0 .63;coefficient of variation was -14% -4 .1% .P=0 .000 and the difference was statistically significant .When TBIL≤342 .1 μmol/L ,r= 0 .994 ;95% CI of HbA1c results with IE‐HPLC method was -0 .09-0 .50;coefficient of variation was -5 .5% -4 .1% .No interference was detectded with the results .When TBIL was 380 .7-549 .3 μmol/L ,negative bias was observed with the results .Conclusion Our data indicated that HbA1c measurement with IE‐HPLC method could resist the interference of hyperlipidemia;When TBIL≤380 .7 μmol/L and HbA1c<16 .3% ,the results could meet the needs of general clinical detection .Clinical staff should choose more specific HbA1c measurement method according to the patient's condition .