Influences of sterile injector on infusion precision of syringe pump
10.7687/J.ISSN1003-8868.2015.09.126
- VernacularTitle:临床用无菌注射器对微量注射泵输液精度影响分析
- Author:
Guitao CHEN
;
Hongwen LIU
- Publication Type:Journal Article
- Keywords:
syringe pump;
disposable sterile injector;
accuracy;
clinical
- From:
Chinese Medical Equipment Journal
2015;(9):126-128
- CountryChina
- Language:Chinese
-
Abstract:
To explore the influences of the sterile injectors with different sources while the same capacity on the infusion precision of the syringe pump. With six injects with two brands were selected as the subjects and an electronic pump for measuring, flow test was carried out with the brands marked on the pump or not. There was significant difference between the flows from the syringe pumps marked with the injector brands and those without the mark of the brands. The sterile injector has to be set before using the syringe pump.