The optimal dose and reaction time of HLA-B27 antibody in flow cytometry
10.3969/j.issn.1673-4130.2015.20.018
- VernacularTitle:流式细胞术检测 HLA-B27最佳抗体剂量与作用时间研究
- Author:
Na LI
;
Wenbo GAO
;
Yuming ZHOU
- Publication Type:Journal Article
- Keywords:
ankylosing spondylitis;
flow cytometry;
human leukocyte antigen B27
- From:
International Journal of Laboratory Medicine
2015;(20):2965-2966
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the optimal dose and reaction time of human leukocyte antigen B27(HLA‐B27)antibody in flow cytometry .Methods Take 52 cases of whole blood in patients with ankyl‐osing spondylitis(AS) .According to HLA‐B27 antibody doses ,samples were divided into two groups:5 μL group and 10 μL group .HLA‐B27‐positive rate were tested after 5 ,10 ,15 min , respectively .Results The HLA‐B27 positive rate of 5 μL group at different reaction time were (84 .16 ± 1 .21)% ,(94 .81 ± 1 .33)% ,(94 .10 ± 1 .26)% ;the positive rate of 10 μL group at different reaction time were (85 .40 ± 1 .27)% ,(96 .76 ± 1 .31)% , (95 .36 ± 1 .45)% .The positive rate of HLA‐B27 in 10 μL group was higher than 10 μL group(F=90 .08 ,P<0 .05) .The positive rate of HLA‐B27 after reacting for 10 and 15 min were higher than that after reacting for 5 min(F=60 .25 ,P<0 .05) .There was not statistically significantly different between the reaction time of 10 min and 15 min(F=1 .08 ,P>0 .05) .Conclusion The opti‐mal dose and reaction time of HLA‐B27 antibody in flow cytometry are 10μL and 10 min;There is not any interaction between anti‐body dose and the reaction time of HLA‐B27 antibody .