The investigation on equipment (SPECT) performance in nuclear medicine
10.3969/J.ISSN.1672-8270.2015.10.003
- VernacularTitle:核医学设备单光子发射计算机成像系统性能现状调查
- Author:
Ying SONG
;
Hui LIU
- Publication Type:Journal Article
- Keywords:
Single-photon emission computed tomography;
Intrinsic uniformity;
Intrinsic spatial linearity;
Performance test
- From:
China Medical Equipment
2015;(10):9-11
- CountryChina
- Language:Chinese
-
Abstract:
Objective:Through 51 SPECT and SPECT/CT systems for performance test results of the collation and analysis, to understand the status of the operation of the SPECT and SPECT/CT systems, and in-depth analysis of indicators and find out the weak links and puts forward suggestions for improvement. Methods:According to the NEMA standards and requirements of the manufacturer, the systems have been tested on intrinsic uniformity, intrinsic spatial linearity, intrinsic spatial resolution, maximum count rate, system sensitivity, system spatial resolution and tomographic spatial resolution. Results:The results show that the majority of SPECT or SPECT/CT systems can pass the test;the failure rate is 9.80%. And the failures mainly show intrinsic uniformity and intrinsic spatial linearity. Conclusion:It is effective to improve the equipment performance by doing well in the daily quality control, regular maintenance and calibration, reducing factors that influence the inherent uniformity and inherent linear.