Development and clinical application effect of venipuncture auxiliary devices under special condi-tions
10.3760/cma.j.issn.1672-7088.2015.25.019
- VernacularTitle:特殊条件下静脉穿刺辅助装置的研制与临床应用效果
- Author:
Lei LIU
;
Ling YE
;
Ping LI
;
Wei SHENG
;
Yanan WANG
- Publication Type:Journal Article
- Keywords:
Infrared rays;
Infusions intravenous;
Nursing care;
Equipment design
- From:
Chinese Journal of Practical Nursing
2015;(25):1935-1937
- CountryChina
- Language:Chinese
-
Abstract:
Objective To design and develop a kind of venipuncture auxiliary device to use in the absence of light condition or patients with poor vein condition, and validate its clinical effect. Methods A total of 70 patients with shock were divided into experimental group (venipuncture auxiliary device group with 37 cases) and control group (non-device group with 42 cases) by random digits table method. The one-time puncture success rate and operating time were compared between two groups. Results The one-time puncture success rate was 94.59%(35/37) in experimental group and 78.57%(33/42) in control group, and there was significant difference, χ2=4.214, P<0.05. The time of establishing a single vein passage was (73.63±29.35) s in experimental group and (107.61±45.94) s in control group, and there was significant difference, t=3.656, P<0.05. Conclusions The device is small in size, low cost, easy to carry, the screen is clear, is a rapid infusion with explore vascular equipment. It can guide nurses accurately on vascular puncture, fast, smooth and effective transfusion puncture process.