Limitation of Staphylococcus Spp Test Kit ATB STAPH 5 from BioMerieux in Clinical Applcation
10.3969/j.issn.1671-7414.2015.03.037
- VernacularTitle:生物梅里埃 ATB STAPH 5葡萄球菌药敏试剂盒的局限性分析
- Author:
Yaxiong XIAO
;
Yusheng PENG
;
Peng WANG
;
Zhongtuan HUANG
;
Xiaoqin DAI
- Publication Type:Journal Article
- Keywords:
biomerieux;
susceptibility;
staphylococcus spp;
limitation
- From:
Journal of Modern Laboratory Medicine
2015;(3):122-125
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the limitation analysis of Staphylococcus spp Susceptibility Test Kit ATB STAPH 5 from Bi-oMerieux in clinical application and to propose remedy.Methods The coverage area and break point of antibiotics ATB STAPH 5 from BioMerieux were analysed by using CLSI M100-S23 as standard.Results The kit ATB STAPH 5 did not include chloromycetin,oxazolidinones and lipopeptide three antibiotics that were recommend by CLSI.The 16 kinds of antibi-otics break point in the kit ATB STAPH 5 were all not inconsistent or did not contain with the standard of CLSI M100-S23. It may lead to error drug sensitivity tests,false sensitive or false drug resistance.Conclusion Only using this kit ATB STAPH 5 for Staphylococcus spp ,it may occurre errors or mistakes.Other experiments must be corrected before the re-port.