Evaluation of the Performance of VITEK-2 Compact,VITEK MS and Bruker MS on the Identification of Corynebacterium
10.3969/j.issn.1671-7414.2014.06.005
- VernacularTitle:Vitek-2 Compact和MALDI TOF MS对棒状杆菌属细菌鉴定能力评估
- Author:
Jingwei CHENG
;
Linying SUN
;
Zhipeng XU
;
Xin HOU
;
Yingchun XU
- Publication Type:Journal Article
- Keywords:
corynebacterium;
vitek-2 compact;
VITEK MS;
bruker MS
- From:
Journal of Modern Laboratory Medicine
2014;(6):15-17
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the performance of VITEK-2 compact,VITEK MS and Bruker MS on the identification of Corynebacterium.Methods This was a methodological evaluation study.The 40 Corynebacterium from bioMerieux were i-dentified with the three methods respectively.16S rDNA gene sequencing was conducted as reference method.Made a de-scriptive analysis of the identification ability,time and cost.Resulets The accuracy of species level of the three methods was 95.0%,88.9% and 97.5%.The mean time was 5~6 h,2~3 min and 2~3 min.The cost of consumable was 50~70 yuan, 15~25 yuan and 10~20 yuan.Conclution Three methods with high accuracy can meet the requirement of clinical diagno-sis,and the identification ability of VITEK MS on Corynebacterium amycolatum need to be further improved.