Study on performance evaluation indicators of central venous pressure measuring system for critical patients in clinical trials
10.3760/cma.j.issn.1672-7088.2015.01.002
- VernacularTitle:急危重症患者中心静脉压测量系统效能评估体系的临床实证研究
- Author:
Lan SHI
;
Xianbin GUO
;
Jinyi HE
;
Hong LI
- Publication Type:Journal Article
- Keywords:
Critical patients;
Central venous pressure;
Performance evaluation;
Nursing
- From:
Chinese Journal of Practical Nursing
2015;31(1):5-8
- CountryChina
- Language:Chinese
-
Abstract:
Objective We aimed to study the performance evaluation indicators of central venous pressure (CVP) measuring system for ICU patients in clinical trials and discuss its impact factors.Methods Ac-cording to the performance evaluation indicators of CVP measuring system based on the Delphi method,a form was created to evaluate the continuous CVP monitoring of 197 ICU patients and then the data were collected.A comparative analysis was performed to verify the factors affecting the performance of CVP measuring system.Results The performance of CVP measuring system was affected by the number of threeway switches,vasoactive drugs and positive end-expiratory pressure,the statistical value was 8.577,-6.773and 3.244.There were no statistically significant differences with respect to performance evaluation indicators of CVP measuring system among the patients with different body positions.Conclusions The performance evaluation indicators of CVP measuring sys-tem proves to be feasible and practical by clinical empirical research.Their influencing factors were dis-cussed and they offer a valuable reference for the determination of the performance of CVP measuring sys-tem in clinical practice.