Determination of squamous cell carcinoma antigen based on the magnetic particles chemiluminescence immunoassay
10.3969/j.issn.1673-4130.2015.13.009
- VernacularTitle:基于磁颗粒化学发光免疫分析方法检测鳞状细胞癌抗原?
- Author:
Feng TIAN
;
Suwen QI
- Publication Type:Journal Article
- Keywords:
squamous cell carcinoma antigen;
N-(aminobutyl)-N-(ethylisoluminol);
chemiluminescence immunoassay
- From:
International Journal of Laboratory Medicine
2015;(13):1823-1825
- CountryChina
- Language:Chinese
-
Abstract:
Objective To establish a rapid and sensitive chemiluminescence immunoassay for detecting squamous cell carcinoma antigen (SCCA)in human serum.Methods Fluorescein isothiocyanate (FITC)and N-(aminobutyl)-N-(ethylisoluminol)(ABEI) were used to label two different monoclonal antibodies of anti-SCCA.SCCA in serum combined with labeled antibodies and formed a sandwiched immunoreaction.After adding the substrate solution,the relative light unit of ABEI was measured.Magnetic particles coated with anti-FITC antibody were used as solid separation carrier.Results The results demonstrated that the method was linear to 22 ng/mL with a detection limit of 0.025 ng/mL.The coefficient of variation (CV)was less than 6% and 7% for intra-assay and inter-assay precision,respectively.Compared with the commercial kit,the proposed method showed a correlation of 0.990 1.Conclu-sion chemiluminescence immunoassay based on magnetic particles displays acceptable performance for quantification of SCCA and is appropriate for use in clinical diagnosis.