Performance test and quality assurance of TruebeamTM
10.3969/J.ISSN.1672-8270.2015.01.016
- VernacularTitle:TruebeamTM直线加速器的性能检测和质量保证
- Author:
Baotian HUANG
- Publication Type:Journal Article
- Keywords:
Linear accelerator;
Performance test;
Quality assurance
- From:
China Medical Equipment
2015;(1):50-53
- CountryChina
- Language:Chinese
-
Abstract:
Objective: To set up the procedures of performance test and quality assurance of TruebeamTM by measuring the mechanical parameters, field parameters, imaging isocenter and respiratory gating systems. Methods:Utilize level and graph paper to test the mechanical parameters; Utilize the film, 3D water tank and dosimeter to measure the corresponding field parameters;Utilize IsocCal phantom to test the shift of imaging isocenter and to use dosimeter to verify the beam on accuracy of respiratory gating systems. Results:All of the tested mechanical parameters were within the tolerance. The coincidence of light field and radiation field was less than 1.2 mm. The flatness was less than 105%except for the flattening filter free beam and the symmetry was less than 102%. The absolute dose error was less than 1%. The shift of imaging isocenter was less than 3 mm and the dose error was within 0.5‰when the respiratory gating function was implemented. Conclusion:TruebeamTM possesses superior mechanical performance and is very suitable to carry out complicated radiotherapy techniques.