Application of ECToolbox software in evaluation of left ventricular ejection fraction
- VernacularTitle:ECToolbox软件计算左心室射血分数的适用性
- Author:
Taisong WANG
;
Jinhua ZHAO
;
Yan XING
- Publication Type:Journal Article
- Keywords:
Left ventricular ejection fraction;
Gated myocardial imaging;
Gated blood-pool imaging
- From:
Chinese Journal of Medical Imaging Technology
2010;26(2):330-332
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the applicability of formula R0, R1, R2 of ECToolbox software in measurement of left ventricular ejection fraction (LVEF) with gated myocardial imaging. Methods Rest gated myocardial SPECT with ~(99m)Tc-MIBI and rest gated blood-pool SPECT with ~(99m)Tc-RBC were performed on 64 patients (44 with CHD, 20 with hypertension). LVEFs were separately calculated with formula R0, R1, R2 of ECToolbox software, and the results were recorded as R0 LVEF, R1 LVEF and R2 LVEF, respectively, and compared with LVEFs from gated blood-pool imaging (GBPI LVEF). Results The LVEFs from formula R0, R1 and R2 were all significantly correlated with GBPI LVEF (r=0.905, 0.905 and 0.903, P<0.05, respectively). The accuracy within ±15% deviation from the GBPI LVEF was 54.30%, 71.40% and 22.90%, and within ±30% was 81.40%, 88.60%,74.30%, respectively. The differences among R0 LVEF, R2 LVEF and GBPI LVEF were statistically significant (P<0.05), while between R1 LVEF and GBPI LVEF was not (P>0.05). Conclusion The formula R1 of ECToolbox software is most applicable in measuring LVEF with gated myocardial imaging.