Application of a puncture guidance supporter in CT intervention
- VernacularTitle:穿刺引导架在CT介入治疗中的应用
- Author:
Weijian FENG
;
Haosheng CHANG
;
Yanjie ZHAO
;
Shu FENG
- Publication Type:Journal Article
- Keywords:
Tomography;
X-ray computed;
Radiology;
interventional;
Puncture guidance supporter
- From:
Chinese Journal of Interventional Imaging and Therapy
2010;7(1):66-69
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the application of a puncture guidance supporter in CT guided biopsy and interventional therapy. Methods Totally 468 patients underwent CT interventional biopsy (n=110) or treatment (n=358) using a puncture guidance supporter. The size of lesions was 2-15 cm. CT scanning was performed to determine the entry point, data of bi-directional needle angle α, θ and depth d before operation. After local anesthesia, the needle was punctured into skin through the puncture guidance supporter, the needle overlap CT positioning laser line in tilt angle (α) was adjusted and the puncture direction at angle θ. After CT scanning confirmed the needle aiming at target lesion, punctures were performed with the needle into the depth to hit the target. Results The total puncture hit rate was 100%, and one step hit rate was 96.79%. The average time from location scanning to hit the target was 8.4 min. The incidence of pneumothorax occurred in thorax management was 1.56%. Conclusion The application of puncture guidance supporter in CT interventional diagnosis and treatment makes operation more accurate, safe and convenient.